Multiparameter Probe AT-SB-PROBE-MP-P-C1

From ATRONIA Wiki

Datasheet

1. Introduction

The SENSBLUE RS‑485 Multiparameter Probe AT‑SB‑PROBE‑MP‑P is an online multiparameter self‑cleaning digital sensor introduced by the company. It adopts an integrated design, and the product is reliable and easy to use. Up to 8 parameters can be measured at the same time, with optional sensor types such as dissolved oxygen, COD, pH, ORP, conductivity/salinity, ammonia nitrogen, turbidity, etc. Using the RS‑485 bus and Modbus/RTU communication protocol, data can be directly transmitted to the acquisition platform.

  The online multiparameter water quality sensor is equipped with an automatic cleaning device that allows setting the automatic cleaning interval and the number of cleaning cycles to match different water quality conditions. The automatic cleaning device can effectively clean the sensor surface, prevent microorganism adhesion, and greatly reduce maintenance costs. Each sensor is equipped with a quick‑plug waterproof connector, making assembly and disassembly easy. The front sensor protection cover prevents damage to the internal sensor. Slots around the protection cover effectively block large suspended particles and organisms from damaging the probe while maintaining measurement accuracy.

2. Features

  • Digital sensor with RS‑485 bus and Modbus/RTU communication protocol.
  • Equipped with an automatic cleaning device that effectively cleans the sensor surface, prevents microorganism attachment, ensures more accurate measurements, and reduces maintenance costs.
  • Optional digital sensors such as dissolved oxygen, COD, conductivity/salinity, turbidity, ammonia nitrogen, pH, ORP, etc., suitable for long‑term online monitoring.
  • Integrated design capable of measuring 8 parameters (including temperature) simultaneously.

3. Technical Specifications

Dissolved Oxygen Probe
Measuring Range 0~20mg/L
Accuracy ±0.4mg/L
Resolution 0.01mg/L
Turbidity Probe
Measuring range and Accuracy 0~100NTU ±3% or ±2NTU
0~1000NTU ±5% or ±3NTU
Resolution 0.1NTU
Conductivity / Salinity Probe
Measuring range 0~5000uS/cm 1uS/cm
Resolution 0~200mS/cm 0.1mS/cm
0~70PSU 0.1PSU
Accuracy ±1.5% F.S.
COD Probe
Measuring range COD Turbidity
0~200mg/L equiv. KHP 0~100NTU
0~500mg/L equiv. KHP 0~200NTU
COD accuracy ±5% F.S.
COD resolution 0.1mg/L
Turbidity accuracy ±5% F.S.
Turbidity resolution 0.1NTU
pH Probe
Measuring range 0~14pH
Accuracy ±0.1pH
Resolution 0.01pH
ORP Probe
Measuring range -1500mV~+1500mV
Accuracy ±10% or ±2mg/L
Resolution 0.1mg/L
Temperature
Measuring range 0~50℃
Accuracy ±0.5℃
Resolution 0.1℃
Other Information of Multi-parameter Probe
Output RS-485 (Modbus/RTU)
Cleaning method Automatic cleaning
Power consumption 5W @ 12V
Power supply 12VDC ±5%
Cable length 5 meters, other lengths can be customized
Chlorophyll

4. Dimensions and Wiring


Note: When the sensor is installed for measurement, use lifting rings or pipe thread installation to avoid direct force on the cable. The sensor connector is M16-5 core waterproof male connector.


Pinouts
1 - NA 4 - RS485 B-
2 - GND 5 – NA
3 - RS485 A+ 6 - Pin +12V DC
7 - NA

5. Maintenance Schedule

SENSBLUE RS-485 Multiparameter Probe AT-SB-PROBEMP-P is equipped with a cleaning brush, which can extend the maintenance cycle. However, due to the diversity of the environment, it is recommended to inspect, clean and calibrate the sensors regularly.


Maintenance Task Recommended Maintenance Frequency
Clean the sensor According to the use environment
Calibrate the sensor (if required by the competent authority) Periodic one- or two-point calibration

5.1 Maintenance Methods

1.Inspection:

  Check whether the sensor´s head is dirty and microorganisms are attached, whether the housing and sensor surface are damaged, whether the cable is normal, whether the test data is normal, and whether the consumables are damaged.

2. Cleaning:

  Clean the outer surface of the sensor with tap water. If debris remains, wipe it with a moist soft cloth. For stubborn dirt, you may add some household detergent to the tap water.

3.Calibration:

  Perform single‑point or two‑point calibration of the sensor. Select the appropriate standard solution according to the corresponding sensor.

5.2 Matters Needing Attention

The probe contains sensitive optical and electronic components. Make sure the probe is not subjected to severe mechanical shock. There are no parts inside the probe that require user maintenance.

6. Modbus Protocol

6.1 Data Format

The default data format for Modbus communication is: 9600, n, 8, 1 (baud rate 9600 bps, 1 start bit, 8 data bits, no parity, 1 stop bit).

  Parameters such as baud rate can be customized.

6.2 Information Frame  Format

a) Read data instruction frame

06                         03                       xx  xx                         xx   xx                    xx  xx            

Address   Function code   Register start address   Number of registers   CRC check code (low byte first)

b) Read data response frame

06                         03                       xx  xx                         xx   xx                   xx  xx

Address    Function code             Bytes                   Answer data     CRC check code (low byte first)

c) Write data instruction frame

06                         06                       xx  xx                         xx   xx                    xx  xx            

Address   Function code   Register address          Write data       CRC check code (low byte first)

d) Write data response frame (same data command frame)

06                         06                       xx  xx                          xx   xx                   xx  xx

Address    Function code    Register ddress            Write data     CRC check code (low byte first)

6.3 Register Address

Register Address Name Instruction Number of Registers Access Method
0x0006 Conductivity / Salinity value 2 double-byte integers: conductivity/salinity value and number of decimal places. 2 (4 bytes) Read (0x03)
0x0008 pH value 2 double-byte integers: pH value and number of decimal places. 2 (4 bytes) Read (0x03)
0x000A ORP value 2 double-byte integers: ORP value and number of decimal places. 2 (4 bytes) Read (0x03)
0x000C DO value 2 double-byte integers: DO value and number of decimal places. 2 (4 bytes) Read (0x03)
0x0010 Turbidity value 2 double-byte integers: turbidity value and number of decimal places. 2 (4 bytes) Read (0x03)
0x0020 Chlorophyll 2×16-bit integers: chlorophyll value and decimal digits. 2 (4 bytes) Read (0x03)
0x1000 Temperature calibration Write: actual temperature ×10. Read: temperature calibration offset ×10. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1002 COD zero calibration Calibration in deionized water. Write value 0 during calibration; read value is the original COD zero-point signal. For 0–200 mg/L range use 0–20 mg/L standard solution; for 0–500 mg/L range use 0–50 mg/L standard solution. Write value = standard solution ×10. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1003 COD slope calibration Calibration in COD standard solutions. Write value = standard solution ×10. Read value = original slope signal. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1004 COD built-in zero turbidity calibration Calibration in deionized water. Write 0; read value = zero offset. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1005 COD built-in turbidity slope calibration Calibration using 20–200 NTU turbidity standard solution. Write = turbidity ×10; read = slope ×1000. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1007 Slope calibration of conductivity / salinity Calibration in standard solution. Write value = actual solution value (or ×10 depending on range). Read value = slope ×1000. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1008 Zero calibration of pH Calibration in pH 6.86 standard solution. Write 0; read value = zero offset. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1009 Slope calibration of pH Calibration in pH 4.00 or pH 9.18 solution. Write 0 (pH4) or 1 (pH9). Read value = slope ×1000. 1 (2 bytes) Write (0x06) / Read (0x03)
0x100A Zero calibration of ORP Calibration in ORP standard solution. Write = actual value; read = zero offset. 1 (2 bytes) Write (0x06) / Read (0x03)
0x100B Slope calibration of ORP Calibration in ORP standard solution. Write = standard value; read = slope ×1000. 1 (2 bytes) Write (0x06) / Read (0x03)
0x100C Zero calibration of DO Calibration in oxygen-free water. Write 0; read value = zero offset. 1 (2 bytes) Write (0x06) / Read (0x03)
0x100E Zero calibration of NH Calibration using 1 ppm or 10 ppm solution. Write value = concentration ×10. Read value = mV ×100 corresponding to zero calibration. 1 (2 bytes) Write (0x06) / Read (0x03)
0x100F Slope calibration of NH+ Calibration using 10 ppm or 100 ppm solution. Write value = concentration ×10. Read value = mV ×100 corresponding to slope calibration. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1010 Zero calibration of turbidity Calibration in DI water or 0–20 NTU solution. Write value = turbidity ×10. Read value = zero offset. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1011 Slope calibration of turbidity Calibration in turbidity standard solution. Write = turbidity ×10; read = slope ×1000. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1020 Chlorophyll zero calibration Calibration in DI water. Write 0; read value = offset. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1021 Chlorophyll slope calibration Calibration using 200–400 µg/L standard solution. Write = value ×10; read = slope ×1000. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1300 Automatic cleaning interval Default 30 minutes. Range: 6–6000 minutes. 1 (2 bytes) Write (0x06) / Read (0x03)
0x1301 Automatic cleaning lap setting Default 3 laps. Range: 0–10 laps. 1 (2 bytes) Write (0x06) / Read (0x03)
0x2000 Temperature sub-sensor address Default value 4. Can be set to 1, 4, 8, 32, 64, 65, etc. 1 (2 bytes) Write (0x06) / Read (0x03)
0x2002 Sensor address Default address 6. Range: 1–127. 1 (2 bytes) Write (0x06) / Read (0x03)
0x2020 Reset Write 0 to restore factory settings for cleaning interval, cleaning cycles, and related parameters. 1 (2 bytes) Write (0x06)