1. Introduction

The SENSBLUE RS-485 Dissolved CO2 Sensor Probe AT-SB-PROBE-CO2-P-5000-C1 is a gas detection module based on NDIR infrared absorption principle, which is suitable for detecting the concentration of carbon dioxide in aqueous solution. KWS-5000 adopts a patented optical cavity, adopts convection diffusion ventilation method and protective cover, which not only accelerates the gas convection diffusion speed, but also waterproof and breathable. Unique waterproof structure, detachable, easy to clean the outer casing of the sensor.

2. Features

  • Unique waterproof structure, can be disassembled;
  • Lightweight housing for easy field installation;
  • Rich interface, convenient for customers to choose;
  • Provides zero calibration, sensitivity calibration and clean air calibration

3. Technical Specifications

Item Parameter
Measuring Range 2000ppm/5000ppm/10000ppm
Response Time <20 sec
Protection Level IP68
Storage Temperature -20 ~ 80°C
Operation Temperature -20 ~ 60°C
Output RS485 / 4-20mA / voltage, IIC, UART, PWM
Power Supply DC 5V (RS485 / 4-20mA is DC12V power supply)
Size Φ27*105mm
Probe Cable Length Default 0.5m
Maximum Working Pressure 1.5 bar
Product Material ABS

4. Dimensions

5. Signal interface definition

Signal definition
SN Wire color UART+IIC interface UART+485 interface UART+4-20mA interface
1 Red 5VDC 12V 12V
2 Black GND GND GND
3 Green Uart_RXD Uart_RXD Uart_RXD
4 Yellow Uart_TXD Uart_TXD Uart_TXD
5 White IIC_SDA RS485A 4-20mA(0.4-2V) output
6 Brown IIC_SCL RS485B Reserved

6. Order information

Tpye Measure range Signal Protocol Cable length Description
KWS5000 Product type
2000 0-2000ppm accuracy±20ppm (default)
5000 0-5000ppm accuracy±50ppm
10000 0-10000ppm accuracy ±300ppm
XXXX customize range
485 RS485 interface
IIC IIC interface
Iout 4-20mA current
Vout 0-5V Vout put
PWM PWM output
RTU (default) Modbus RTU Protocol
ACSII Modbus ASCII Protocol
self defined customized protocol
0.5m (default) customized
KWS5000 2000 485 RTU 0.5 Example configuration

7. Applicable Gateways & Accessories

Name Requirements
PA000000076 SENSBLUE MONARCH Solar Panel 1 Probe
PA000000079 SENSBLUE MONARCH Solar Panel 2 Probes
PA000000080 SENSBLUE MONARCH Solar Panel 1 Probe External Antenna
PA000000074 SENSBLUE MONARCH Solar Panel 2 Probes External Antenna
PA000000081 SENSBLUE MONARCH Opaque Case 1 Probe
PA000000082 SENSBLUE MONARCH Opaque Case 2 Probes
PA000000083 SENSBLUE MONARCH Opaque Case 1 Probe External Antenna
PA000000084 SENSBLUE MONARCH Opaque Case 2 Probes External Antenna
PA000000020 SENSBLUE MONARCH RS485 Probe Splitter Box (1 input -> 2 outputs)

8. Other Probes

Name Firmware v. Requirements
PA000000065 SENSBLUE RS-485 Dissolved Oxygen Probe AT-SB-PROBE-DO-P-C1, w/ 15m cable w/connector
PA000000085 SENSBLUE RS-485 Dissolved Oxygen Probe AT-SB-PROBE-DO-T-C1, w/ 15m cable w/connector 10.70
PA000000091 SENSBLUE RS-485 Dissolved CO2 Sensor Probe AT-SB-PROBE-CO2-P-5000-C1, w/ 15m cable w/connector
PA000000086 SENSBLUE RS-485 Low Conductivity Probe AT-SB-PROBE-LC-P-C1, w/ 15m cable w/connector 10.70
PA000000087 SENSBLUE RS-485 Medium Conductivity Probe AT-SB-PROBE-MC-P-C1, w/ 15m cable w/connector 10.70
PA000000068 SENSBLUE RS-485 High Conductivity Probe AT-SB-PROBE-HC-P-C1, w/ 15m cable w/connector 10.70
PA000000089 SENSBLUE RS-485 Salinity/Conductivity/TDS Probe AT-SB-PROBE-S-P-C1, w/ 15m cable w/connector
PA000000067 SENSBLUE RS-485 pH Probe AT-SB-PROBE-pH-P, w/ 15m cable w/conector 10.70
PA000000088 SENSBLUE RS-485 ORP Probe POM AT-SB-PROBE-ORP-P, w/ 15m cable w/conector
PA000000069 SENSBLUE RS-485 Nitrates (NO3) Probe AT-SB-PROBE-NO3-P, w/ 15m cable w/connector 10.70
PA000000090 SENSBLUE RS-485 Ammonia Nitrogen (NH4-N) Sensor Probe AT-SB-PROBE-NH4-P-C1, w/ 15m cable w/connector
PA000000066 SENSBLUE RS-485 Turbidity Probe AT-SB-PROBE-T-P, w/ 15m cable w/connector 10.70
PA000000073 SENSBLUE RS-485 Multiparameter Probe AT-SB-PROBE-MP-P-C1, w/ 15m cable w/connector 10.71

9. Applicable certifications and standards

EMC Directive 2014/30/EU.

  • EN IEC 61326-1: 2021
  • EN 55011: 2016+A2: 2021
  • EN IEC 61000-3-2: 2019+A1: 2021
  • EN 61000-3-3: 2013+A2: 2021