RS-485 ORP Probe POM AT-SB-PROBE-ORP-P

From ATRONIA Wiki
Revision as of 11:48, 13 July 2026 by WikiManagement (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

Datasheet

1. Introduction

SENSBLUE Monarch ORP Probe POM AT-SB-PROBE-ORP-P can be used for environmental water quality monitoring, acid/alkali/salt solution, chemical reaction process, industrial production process, and can meet the requirements of online ORP measurement for most industrial applications.

2. Features

  • Signal output: RS-485 (Modbus/RTU protocol).
  • Conveniently connect to third-party equipment such as PLC, DCS, industrial control computer, general controller, paperless recording instrument or touch screen.
  • Double high impedance differential amplifier, strong anti-interference and fast response speed.
  • Durable ORP electrode, the internal reference fluid seeps out from the microporous salt bridge very slowly under a pressure of at least 100KPa (1Bar), and its positive seepage lasts for more than 20 months. Such a reference system is very stable, and the electrode life is doubled compared to ordinary industrial electrodes.
  • Easy to install: 3/4NPT pipe thread is convenient for sinking installation or installation in pipelines and tanks.
  • IP68protection level.
  • R-485 interface, Modbus/RTU protocol.

3. Technical Specifications

Model AT-SB-PROBE-ORP-P
Measuring Range -1500~+1500mV
Accuracy ±6mV
Resolution 1mV
Temperature Range 0 ~65℃
Work Pressure <0.1MPa
Output RS-485, MODBUS protocol
Power Supply 12~24VDC ±10%
Wetted Part Material POM
Installation Immersion mounting, 3/4 NPT thread
Cable Length 15m
Calibration Method One-point calibration
Power Consumption <0.3W@12V
Ip Grade IP68

4. Dimension

5. Installation and Electrical Connection

5.1 Install

Note: The sensor should not be installed upside down or horizontally when installed, at least at an angle of 15 degrees or more.

5.2 Electrical connection

  • Red wire-power cord(12~24 V)
  • Black wire-ground wire(GND)
  • Blue Line-485 A
  • White Line-485 B

Check the wiring sequence carefully before powering on to avoid unnecessary losses caused by wiring errors.

5.3 Wiring instructions

Considering that the cables have been immersed in water (including seawater) or exposed to air for a long time, all wiring positions are required to be waterproof, and the cables should have certain corrosion resistance.

6. Maintenance

6.1 Use and maintenance

When measuring the ORP electrode, it should be cleaned in distilled water (or deionized water) first, and the water should be absorbed with filter paper to prevent impurities from being brought into the test solution. 1/3 of the electrode should be inserted into the test solution. The electrode should be cleaned when not in use and inserted into a protective cover with 3.5mol/L potassium chloride solution or inserted into a container with 3.5mol/L potassium chloride solution. Check whether the terminal is dry, if it is stained, wipe it with absolute alcohol and blow it dry before use. Avoid long-term immersion in distilled water or protein solution and prevent contact with silicone grease. For electrodes that have been used for a long time, its glass membrane may become translucent or attached with sediment. In this case, it can be washed with diluted hydrochloric acid and rinsed with water. The electrode has been used for a long time, and when there is a measurement error, it must be calibrated and corrected with the meter. When the electrode is still unable to be calibrated and measured when the electrode is maintained and maintained by the above method, it means that the electrode has failed. Please replace the electrode.

6.2 ORP standard solution preparation method

a) Preparation method of ORP standard solution 256mV Weigh 2.1 g of quinhydroquinone and add it to 200 mL of a standard pH buffer solution with a pH of 4.00 and mix well.

b) Preparation method of ORP standard solution 86mV Weigh 2.1 g of quinhydroquinone and add it to 200 mL of a standard pH buffer solution with a pH of 6.86 and mix well.

c) Preparation method of ORP standard solution-40mV Weigh 2.1 g of quinhydroquinone and add it to 200 mL of a standard pH buffer solution with a pH of 9.18 and mix well.

Please note that when preparing the pH, a total of 250 mL is prepared, and 200 mL is measured for use.

Method of preparing corresponding standard solution for pH:

Measure 250mL of distilled water with a graduated cylinder, pour it into a beaker, add a package of calibration powder with pH=6.86/4.00/9.18, stir with a glass rod until the powder is completely dissolved, and prepare a solution with pH=6.86/4.00/9.18.

6.3 Calibration

Put the sensor into the prepared 86mV (or 256mV or-40mV) solution and wait for 3 to 5 minutes. After the value is stable, check whether the displayed value is 86mV (or 256mV or-40mV). If not, you need to calibrate it. See the appendix for calibration instructions.

6.4 Precautions

  • Avoid exposing the inner surface of the fluorescent film head to sunlight
  • Please do not touch the fluorescent film with your hands
  • Avoid bubbles adhering to the surface of the fluorescent film during measurement and calibration
  • Avoid applying any mechanical stress (pressure, scratches, etc.) directly to the fluorescent film during use

7. Appendix data communication

7.1 Data Format

The default data format for Modbus communication is: 9600, n, 8, 1 (baud rate 9600bps, 1 start bit, 8 data bits, no parity, 1 stop bit).

7.2 Modbus Frame Format (xx represents a byte)

a) Read data instruction frame

06 03 xx xx xx xx xx xx
Address FC Register start address Number of registers CRC check code (low bytes in front)

b) Read data response frame

06 03 xx xx xx xx xx xx
Address FC Number of bytes Response data CRC check code (low bytes in front)

c) Write data instruction frame

06 06 xx xx xx xx xx xx
Address FC Register address Read-in data CRC check code (low bytes in front)

d) Data response frame

06 06 xx xx xx xx xx xx
Address FC Register address Read-in data CRC check code (low bytes in front)

7.3 Register Address

Register Address Name Instruction Number of Registers Access Method
40001 (0x0000) Measured value Two double-byte integers are the measured value and the number of decimal places of the measured value. 2 (4 bytes) Read
44097 (0x1000) Zero calibration Calibrate in the standard solution, the written data is the actual value of the standard solution. 1 (2 bytes) Write
44103 (0x1006) Zero calibration value The read data is the zero offset. 1 (2 bytes) Read
48195 (0x2002) Sensor address The default is 6, and the write data range is 1-127. 1 (2 bytes) Write/Read
48225 (0x2020) Reset sensor The calibration value is restored to the default value, and the write data is 0. Note that the sensor needs to be calibrated again after reset before it can be used. 1 (2 bytes) Write
44105(0x1008) Slope calibration value The read data is the slope value x1000 1 ( 2 bytes) Read
44113(0x1010) Temperature calibration Calibrated in solution, the data written is the actual temperature value x10 ; the read data is the temperature calibration  offset x10 . 1 ( 2 bytes) Write/Read
48195(0x2002) Sensor address The default is 6 and the data range is 1-127 1 ( 2 bytes) Write/Read
48225(0x2020) Reset sensor The calibration value is restored to the default value and the write data is 0. Note: the sensor needs to be calibrated again after resetting. 1 ( 2 bytes) Write

Note:

a) The register address is the starting address of the register with register type defined according to the Modbus protocol (the actual starting address of the register represented by the hexadecimal system in parentheses).

b) When changing the sensor address, the sensor address in the return instruction is the new address after the change.

c)The data definition of the measured value returned when reading the data:

xx xx xx xx

2-byte measurement value 2-byte measurement value decimal places

The default data type is: double-byte integer, high byte first; other types such as floating-point numbers are optional

7.4 Command example

a) Start measurement command

Function: Obtain the ORP measured by the sensor, the unit of ORP is mV.

Request frame: 06 03 00 00 00 02 C5 BC

Response frame: 06 03 04 00 78 00 00 0C EA

Examples of readings:

ORP value
00 78 00 00

For example, ORP value: 00 78 represents the hexadecimal reading ORP value, 00 00 represents the ORP value without a decimal point and no temperature value, converted to a decimal value of 120.

b) Calibration instruction:

The ORP sensor only needs one point calibration. Before calibrating with another calibration solution, the electrode must be cleaned with the latter calibration solution, otherwise it will cause a larger test error.

In the prepared 86mV or 256mV standard solution, enter the corresponding calibration instructions.


The standard solution is 86mV:

Request frame: 06 06 10 00 00 56 0C 83;

Response frame: 06 06 10 00 00 56 0C 83;


The standard solution is 256mV:

Request frame: 06 06 10 00 01 00 8D 2D;

Response frame: 06 06 10 00 01 00 8D 2D

Note: The ORP value has a negative value, such as-40mV, its calculation method is:-{FFFF-FFD7 software test value)}=-28; decimal is-40mV.


c) Set the device ID address:

Function: set the MODBUS device address of the electrode.

Change the device address 06 to 01, with the following example:

Request frame:06 06 20 02 00 01 E3 BD

Response frame:06 06 20 02 00 01 E3 BD

7.5 Error response

If the sensor does not execute the upper computer command correctly, the following format information is returned:

Definition Address FC CODE CRC check
data ADDR COM+80H xx CRC 16
Number of bytes 1 1 1 2

a) CODE: 01 – Functional code error

03 – Data error

b) COM: Received function code

8. Applicable Gateways & Accessories

Name Requirements
PA000000076 SENSBLUE MONARCH Solar Panel 1 Probe
PA000000079 SENSBLUE MONARCH Solar Panel 2 Probes
PA000000080 SENSBLUE MONARCH Solar Panel 1 Probe External Antenna
PA000000074 SENSBLUE MONARCH Solar Panel 2 Probes External Antenna
PA000000081 SENSBLUE MONARCH Opaque Case 1 Probe
PA000000082 SENSBLUE MONARCH Opaque Case 2 Probes
PA000000083 SENSBLUE MONARCH Opaque Case 1 Probe External Antenna
PA000000084 SENSBLUE MONARCH Opaque Case 2 Probes External Antenna
PA000000020 SENSBLUE MONARCH RS485 Probe Splitter Box (1 input -> 2 outputs)

9. Other Probes

Name Firmware v. Requirements
PA000000065 SENSBLUE RS-485 Dissolved Oxygen Probe AT-SB-PROBE-DO-P-C1, w/ 15m cable w/connector
PA000000085 SENSBLUE RS-485 Dissolved Oxygen Probe AT-SB-PROBE-DO-T-C1, w/ 15m cable w/connector 10.70
PA000000091 SENSBLUE RS-485 Dissolved CO2 Sensor Probe AT-SB-PROBE-CO2-P-5000-C1, w/ 15m cable w/connector
PA000000086 SENSBLUE RS-485 Low Conductivity Probe AT-SB-PROBE-LC-P-C1, w/ 15m cable w/connector 10.70
PA000000087 SENSBLUE RS-485 Medium Conductivity Probe AT-SB-PROBE-MC-P-C1, w/ 15m cable w/connector 10.70
PA000000068 SENSBLUE RS-485 High Conductivity Probe AT-SB-PROBE-HC-P-C1, w/ 15m cable w/connector 10.70
PA000000089 SENSBLUE RS-485 Salinity/Conductivity/TDS Probe AT-SB-PROBE-S-P-C1, w/ 15m cable w/connector
PA000000067 SENSBLUE RS-485 pH Probe AT-SB-PROBE-pH-P, w/ 15m cable w/conector 10.70
PA000000088 SENSBLUE RS-485 ORP Probe POM AT-SB-PROBE-ORP-P, w/ 15m cable w/conector
PA000000069 SENSBLUE RS-485 Nitrates (NO3) Probe AT-SB-PROBE-NO3-P, w/ 15m cable w/connector 10.70
PA000000090 SENSBLUE RS-485 Ammonia Nitrogen (NH4-N) Sensor Probe AT-SB-PROBE-NH4-P-C1, w/ 15m cable w/connector
PA000000066 SENSBLUE RS-485 Turbidity Probe AT-SB-PROBE-T-P, w/ 15m cable w/connector 10.70
PA000000073 SENSBLUE RS-485 Multiparameter Probe AT-SB-PROBE-MP-P-C1, w/ 15m cable w/connector 10.71

10. Applicable certifications and standards

EMC Directive 2014/30/EU.

  • EN IEC 61326-1: 2021
  • EN 55011: 2016+A2: 2021
  • EN IEC 61000-3-2: 2019+A1: 2021
  • EN 61000-3-3: 2013+A2: 2021