Salinity Conductivity TDS Probe AT-SB-PROBE-S-P-C1
Datasheet
1. Introduction
"The SENSBLUE RS-485 Salinity Conductivity TDS Probe (AT-SB-PROBE-S-P-C1) introduces a new generation of four-electrode conductivity sensor, adopts international leading four-electrode technology, RS485 digital interface, supports MODBUS protocol, and environmentally friendly design. Compared with the traditional two-electrode conductivity sensor, it has higher accuracy, wider measurement range and excellent stability. The four-electrode conductivity sensor also has a unique advantage in quantity: one is to completely solve the polarization problem in high conductivity test; the other is to solve the problem of inaccurate reading caused by electrode contamination.
2. Features
- Simultaneous output of conductivity, salinity, TDS, and temperature parameters;
- Corrosion-resistant housing with IP68 waterproof rating for long-term underwater operation;
- Built-in temperature sensor with automatic temperature compensation;
- Factory-calibrated with built-in calibration parameters for immediate installation and use
3. Technical Specifications
| Model | KWS-300A | KWS-300B |
| Conductivity Range | 0-200mS/cm (0-2mS/cm, 2-20mS/cm, 20-200mS/cm) | |
| Accuracy | ±1% or 0.01mS/cm (which is bigger) | |
| Salinity Range | 0 -175 ppt | |
| Accuracy | ±1 ppt | |
| TDS Range | 0 – 128000 mg/L | |
| Response Time | <10s | |
| Ip Grade | IP68 | |
| Max. Pressure | 6 bar | |
| Sheath Material | Titanium | |
| Operating Temperature | 0~50℃ (no freezing) | |
| Output | RS485, Modbus | |
| Power Consumption | 0.2W (Recommended power supply: DC 9–24V, >500mA) | |
| Sensor Dimension | φ22mm*175.5mm | φ22mm*165.8mm |
| Assembly | NPT3/4 male thread | Connect with multi-parameter host |
| Cable Length | 10 meters (default), customizable | None |
| Calibration | support one point or two-points calibration | |
| Casing Material | Titanium + PEEK | |
Note:
The above technical specifications are based on laboratory standard solution conditions;
Sensor lifespan and calibration/maintenance frequency depend on actual site conditions.