RS-485 Turbidity Probe AT-SB-PROBE-T-P

From ATRONIA Wiki
Revision as of 15:23, 25 March 2026 by WikiManagement (talk | contribs)

Datasheet

1. Introduction

SENSBLUE RS-485 Turbidity Probe AT-SB-PROBE-T-P is used in environmental water quality monitoring, acid/alkali/salt solutions, chemical reaction processes, and industrial production processes. It can meet the requirements of online pH measurement for most industrial applications.

2. Features

  • 90° angle scattered light principle, built-in temperature sensor
  • Support RS-485, Modbus/RTU protocol
  • Optical fiber structure, strong resistance to external light interference
  • Infrared LED light source with high stability
  • IP68 protection, water depth within 20 meters
  • Convenient, fast, stable and easy to maintain

3. Technical Specifications

Model KWS-960
Measuring Range Turbidity PPR AT-SB-PROBE-T-P
Principle Light scattering method
Range/Resolution 0~20.00NTU 0.01NTU/0.1℃
0~200.0NTU 0.01NTU/0.1℃
0~1000.0NTU 0.1NTU/0.1℃
Precision ±3% or ±1.5NTU(0~20.00NTU)
±3% or ±2NTU(0~200.0NTU)
±5% or ±3NTU(0~1000.0NTU)
Response Time (T90) <60s
Lowest Detection Limit 0.01 (0-20NTU)
0.3NTU
Calibration Mode Two-point calibration
Temperature Compensation Automatic temperature compensation (Pt1000)
Output RS-485 (Modbus/RTU)
Working Conditions 0~50℃,<0.2MPa
Storage Temperature - 5 ~ 65 ℃
Installation Immersion mount with 3/4NPT pipe thread
Cable Length 5 meters, other lengths can be customized
Power Consumption 0.2W@12V
Power Supply 12~24VDC
Protection Grade IP68

4. Dimensions

 

Note: The sensor joint is M16-5 core waterproof joint male head

5. Installation and Pinouts

5.1 Installation

Installation distance requirements: keep above 5cm with side wall and above 10cm with bottom.

 
Pinouts
1 - Pin 12-24V DC 4 - RS485 B-
2 - GND 5 – NA
3 - RS485 A+ 6 - NA
7 - NA

6. Maintenance

6.1 Use and maintenance

a) Maintenance schedule

The cleanliness of the measuring window is very important for maintaining accurate readings.

Maintenance task Recommended maintenance frequency
Calibrate sensors

(if required, by the competent authority)

According to the maintenance schedule required by the competent department